Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 3 de 3
Filter
Add filters








Language
Year range
1.
Korean Journal of Medical Physics ; : 29-34, 2010.
Article in Korean | WPRIM | ID: wpr-30107

ABSTRACT

In this study, we have fabricated a fiber-optic dosimeter for a proton beam therapy dosimetry. We have measured scintillating lights with the various kinds of organic scintillators and selected the BCF-12 as a sensor-tip material due to its highest light output and peak/plateau ratio. To determine the optimum diameter of BCF-12, we have measured scintillating lights according to the energy losses of proton beams in a water phantom. Also, we determined the adequate length of organic scintillator by measuring scintillating lights according to the incident angles of proton beam. Using an optimized fiber-optic dosimeter, we have measured scintillating lights according to the dose rates and monitor units of proton accelerator.


Subject(s)
Light , Optical Fibers , Organothiophosphorus Compounds , Proton Therapy , Protons , Water
2.
Journal of the Korean Academy of Family Medicine ; : 22-29, 1991.
Article in Korean | WPRIM | ID: wpr-93523

ABSTRACT

No abstract available.


Subject(s)
Suicide, Attempted
3.
Journal of the Korean Academy of Family Medicine ; : 36-46, 1991.
Article in Korean | WPRIM | ID: wpr-93521

ABSTRACT

No abstract available.

SELECTION OF CITATIONS
SEARCH DETAIL